Catalog Number 1CX*FX25E
Medical Device Problem Code Failure to Prime (1492)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D4: udi - the udi is not required for the reported product code/lot number combination, however the di portion was provided.The actual device has not been returned to the manufacturing facility for evaluation.For this reason, evaluation code 11 was used in the conclusions section of h6.A follow up report will be submitted within 30 days of this report being sent.A review of the device history record and the product release decision control sheet of the involved product/lot # combination was conducted with no relevant findings.A search of the complaint file did not find any other report with the involved product code/lot# combination.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.9681834.Exemption number e2015022.All available information has been placed on file in quality assurance for appropriate tracking, trending and follow-up.
 
Event or Problem Description
The user facility reported that the capiox device failed to prime during setup.Follow up communication with the user facility confirmed the following information; the locked volume out of the bottom vent would not prime; it would prime during normosol; the product was changed out; the procedure was completed successfully; and there was no patient involvement.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide additional information about the device return in device available for evaluation?.The actual device is not available for evaluation.A follow up report will be submitted once the investigation is complete.For this reason, (b)(4) is referenced.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.2 to provide the retention sample evaluation results, provide the entire udi no.& the expiration date.The actual device was not returned to the manufacturing facility.Therefore, the investigation was based upon evaluation of user facility information and a retention sample of the involved product code/lot number combination.Visual inspection of the retention sample revealed no defects.Functional testing was conducted and could not duplicate the reported event.The investigation verified that the retention sample was the normal product.During the inspection, the reported event was not duplicated.With the absence of the actual sample to be evaluated, the cause of this complaint cannot be determined.All available information has been placed on file in quality assurance at the manufacturing facility for appropriate tracking, trending and follow-up.
 
Additional Manufacturer Narrative
This report is being submitted as follow up number 3 to provide the returned sample evaluation results.The actual device was returned to the manufacturing facility for evaluation.Visual inspection found no anomalies.The actual sample was filled with saline solution.With the blood outlet port clamped and pressure applied to the blood pathway from the blood inlet port side for 6 hours.No leak was confirmed.The water pathway of the heat exchanger module was filled with water.With the water outlet port clamped pressure was applied to the water pathway of the heat exchanger module from the water inlet port side for 6 hours.No leak was confirmed.The actual sample was built into a circuit with tubes.Saline solution colored for better visibility was let to flow into the circuit little by little.As a result, the reported complaint could not be duplicated.There is no evidence that this event was related to a device defect or malfunction.The actual sample was confirmed to be normal product.The exact cause of the reported event cannot be definitively determined based on the available information.All available information has been placed on file in quality assurance at the manufacturing facility for appropriate tracking, trending and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX FX25 HOLLOW FIBER OXYGENATOR AF
Common Device Name
OXYGENATOR
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, shizuoka 418
JA  418
MDR Report Key6310339
Report Number9681834-2017-00007
Device Sequence Number16151965
Product Code DTZ
UDI-Device Identifier04987350701060
UDI-Public04987350701060
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K071494
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2019
Device Catalogue Number1CX*FX25E
Device Lot Number160721
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/13/2017
Supplement Date Received by ManufacturerNot provided
Not provided
Not provided
Initial Report FDA Received Date02/08/2017
Supplement Report FDA Received Date03/10/2017
03/28/2017
04/25/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-