| Model Number |
427 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/27/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device has not been returned to numed for evaluation.From the report given to numed by the user facility, it stated that the physician inserted the device and then removed from the body to reposition.There is a specific warning in the instructions for use that states; " pulling the covered stent back through the introducer and/or hemostasis valve may cause the covering to catch and tear off of the stent." this is likely what happened in this instance to cause the covering to loosen from the stent.
|
| |
|
Event or Problem Description
|
|
As per the report from (b)(6): "physician went into the body with the unmounted stent on a bib balloon, and removed the stent from the body to reposition.It was then noticed that the covering was starting to come off the stent.Physician decided not to re-enter the body with this stent, then opened a pre mounted cp stent in the same size which was deployed without incident.".
|
| |
|