Catalog Number 1012452-08
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/02/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event or Problem Description
It was reported that during preparation a 3.75 x 8 mm trek rx balloon dilatation catheter (bdc) was removed without resistance from the dispenser coil and a kink was noted in the shaft.The physician attempted to straighten the catheter.The device was not used and there was no patient involvement.Another device was used to continue the procedure.There was no clinically significant delay in the procedure.No additional information was provided.Returned device revealed the proximal balloon marker was separated.
 
Additional Manufacturer Narrative
(b)(4).Evaluation summary: the device returned for analysis.The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.On (b)(6) 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda on (b)(6) 2017 [medwatch # 2024168-2017-02310].Corrective action has been implemented per site operating procedures.The product will continue to be trended.The abbott internal recall number is (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6311898
Report Number2024168-2017-01037
Device Sequence Number10954723
Product Code LOX
UDI-Device Identifier08717648152016
UDI-Public(01)08717648152016(17)190531(10)60603G1
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K110134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Remedial Action Recall,Other
Type of Report Initial,Followup
Report Date (Section B) 04/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Catalogue Number1012452-08
Device Lot Number60603G1
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/21/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/16/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/08/2017
Supplement Report FDA Received Date04/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-