Catalog Number IGTCFS-65-2-UNI-CELECT-PT
Device Problem Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/04/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Catalog#: igtcfs-65-2-uni-celect-pt.(b)(4).Summary of investigational findings: unfortunately, no product was returned and without the actual complaint device, it would be inappropriate to speculate at what may or may not have occurred based on the limited information provided and it cannot be determined why the filter "would not advance fully through haemostatic valve" in the first place.However, since the filter was damaged during removal, it is suggested that the filter legs were damaged when attempts were made to withdraw the femoral loaded filter through the haemostatic valve.No evidence to suggest product was not manufactured according to specifications.Cook medical will continue to monitor for similar events.
 
Event Description
Description of event according to initial reporter: product would not advance fully through hemostatic valve so removed and filter became damaged right femoral access gianes, flushed filter introducer positioned and venocavagram taken.Filter on femoral set put into introducer but would not advance fully through heamostatic valve so removed and filter became damaged.New filter set opened and placed with no further issue.Patient outcome: the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
4632 bjaeverskov
DA 
Manufacturer Contact
marianne hoey
sandet 6
bjaeverskov 4632
DA   4632
56868686
MDR Report Key6313896
MDR Text Key67249844
Report Number3002808486-2017-00347
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002345048
UDI-Public(01)10827002345048(17)190519(10)E3451048
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
K121629
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-2-UNI-CELECT-PT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2016
Initial Date FDA Received02/09/2017
Date Device Manufactured05/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-