MAUDE Adverse Event Report: SYNTHES OBERDORF DOUBLEAIRHOSE L3M F/SYST-SYNTHES; INSTRUMENT SURGICAL, ORTHO, PNEUMATIC, POWERED AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES OBERDORF DOUBLEAIRHOSE L3M F/SYST-SYNTHES; INSTRUMENT SURGICAL, ORTHO, PNEUMATIC, POWERED AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 519.510
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgery, it was observed that the air hose device was leaking air from the hose.It was not reported if there was a delay to the surgical procedure or whether a spare device was available for use.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device, observed that the device passed all manufacturing specifications and identified no failure.Therefore, the reported condition was not duplicated and confirmed.An assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOUBLEAIRHOSE L3M F/SYST-SYNTHES
Type of Device
INSTRUMENT SURGICAL, ORTHO, PNEUMATIC, POWERED AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ   CH4436
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6314140
MDR Text Key67210370
Report Number8030965-2017-10693
Device Sequence Number1
Product Code HSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number519.510
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/23/2017
Initial Date FDA Received02/09/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-