|
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, a bend with twisted shaft material was identified approximately 10cm from the distal tip, which is distal to the transition point.A kink was identified 14cm from the distal tip, which is distal to the transition point.The stiffening bar on the interior of the catheter sheathing appeared to be bent and or twisted distal to the transition point.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The most likely root cause of the kinks distal to the transition point of the catheter is improper manipulation of the catheter during the procedure and/or excessive pressure exerted during navigation and manipulation of the catheter.The instructions for use (ifu) state: do not introduce the tip folded into the guiding sheath.Do not exert excessive pressure during placement of catheter if unknown resistance is encountered.Do not attempt to use the reprocessed ep catheter prior to completely reading and understanding the directions for use.Avoid manual pre-bending of distal curve, as this may damage steering mechanism of steerable catheters.The reported event will continue to be monitored through post-market surveillance.
|