• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 174301
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection, the returned complaint device showed signs of clinical use.The blades had debris present.Inspection of the blades revealed wear.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause(s) are: - user attempts to cut staples, clips, or non-soft tissue - wear to blades due to clinical use the instructions for use (ifu) state: - do not directly apply current to staples or clips.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported that the scissors were too dull to use during a lap chole.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
MDR Report Key6316001
Report Number0001056128-2017-00027
Device Sequence Number1610143
Product Code NUJ
UDI-Device Identifier00885825012564
UDI-Public(01)00885825012564(17)190610(10)4650648
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K012700
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/10/2019
Device Model Number174301
Device Catalogue Number174301CO
Device Lot Number4650648
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2017
Initial Report FDA Received Date02/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-