| Catalog Number |
07H77-01 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/16/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer stated that the cd1800 analyzer generated an elevated wbc result of 62,300/ul.The sample was repeated on another cd1800 in the lab and a result of 9,900/ul was generated.The customer performed an autoclean on the original analyzer and repeated the sample and a result of 10,000/ul was generated.There was no report of impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
A field service representative (fsr) visitied the account and found that the wbc transducer was damaged.The fsr replaced the transducer to resolve the issue.Customer complaint data was reviewed and no adverse trends were identified.The cell-dyn 1800 system operations manual was reviewed and was found to adequately address the issue.The investigation did not identify a product deficiency.
|
| |
|