Model Number 229047
Device Problem Failure to Capture (1081)
Patient Problem Non specific EKG/ECG Changes (1817)
Event Type  Injury  
Manufacturer Narrative
Product event summary: at analysis the device was noted to be mechanically intact and passed all incoming functional tests.The battery voltage was low (depleted) though this was additionally noted to be normal.The battery was replaced and the device passed functional and systems tests.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the software analyzer experienced a loss of capture.It was further reported via follow-up that this occurred while threshold tests were being performed during an implant.The analyzer was changed out for another and the procedure was completed however it was noted that the patient experienced some pauses but no long-term consequences.The analyzer was returned for service.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6318382
MDR Text Key67084946
Report Number2182208-2017-00189
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number229047
Device Catalogue Number229047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2016
Date Manufacturer Received11/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
-
-