Catalog Number 8065751912
Device Problem Leak/Splash (1354)
Patient Problem No Code Available (3191)
Event Type  Malfunction  
Manufacturer Narrative

No lot number was identified with this complaint; therefore, lot history and complaint history reviews could not be conducted. No sample has been returned for evaluation; therefore, the condition of the product could not be verified. Due to an observed increased trend for this event, an internal investigation was initiated to determine if corrective and preventative actions were necessary. A root cause for the increased complaint rate was found to be related to a manufacturing post assembly inspection practice. This complaint does not identify a reported lot and no sample was returned for evaluation so it cannot be determined if the complaint can be attributed to the post assembly inspection. The post assembly inspection has been discontinued and an effectiveness check has been established and will be monitored periodically. (b)(4).

 
Event Description

An ophthalmic surgeon reported that the valved trocars leaked and produced intraocular pressure instability during half of the procedures performed. Additional information has been requested. The product samples were discarded by the facility.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCONSTELLATION SURGICAL PROCEDURE PAK
Type of DeviceGENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key6321664
MDR Text Key67380190
Report Number1644019-2017-00140
Device Sequence Number1
Product Code LRO
Combination Product (Y/N)N
Reporter Country CodeSP
PMA/PMN NumberK880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation HEALTH PROFESSIONAL
Type of Report Initial
Report Date 02/10/2017
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received02/10/2017
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue Number8065751912
Device LOT NumberASKU
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/27/2017
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device?
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 02/10/2017 Patient Sequence Number: 1
-
-