|
This is initial mdr report being submitted for this complaint with associated mfr# 2954740-2017-00028.(b)(4).Concomitant medical products: guidewire (synchro, stryker/traccess14/transend, stryker); guiding catheter (5f fubuki dilator, asahi intecc); micro catheter (sl-10, stryker); balloon catheter (tranceform 4*10, stryker); enpower.The product is available for evaluation and testing, however the analysis has not been completed.Additional information will be submitted within 30 days of receipt.No conclusions are made at this time.
|
|
This is final mdr report being submitted for this complaint with associated mfr# 2954740-2017-00028.Limited information was received.The unidentified detachment control box (dcb) and the connecting cable were not returned.The sl-10 and the balloon microcatheters along with the rotating hemostatic valve (rhv) were not returned.Concerning cleanliness only, the microcoil system was returned in almost pristine condition.The system was either not used or was cleaned before being returned which may have produced further damage.It is also unknown if the device was improperly handled for returned packaging or was further manipulated and/or inspected post-procedurally.In addition, any trace or other evidence that may have been complaint related may have been altered or removed prior to being returned due to post-procedural handling, cleaning, and packaging.All location and measurement callouts are approximate and for reference only.As viewed though the returned packaging, unreported coil detachment was found.The detached coil was found wrapped around the sheath.The distal dpu of the dpu exhibits multiple detachment attempts.The detachment fiber was found adhering to the melted detachment fiber with blood intermingled.As viewed through the returned packaging, unreported kinks were found on the device positioning unit (dpu) located off the proximal end at 60.7 centimeters and 166.5 centimeters.The device positioning unit (dpu) passed electrical testing with resistance at 50.8 ohms (range 48.5/56.0) and the test enpower and cable systems ready green light illuminated (deltaplush ifu).No manufacturing defects were found.The circumstances of how and when all the unreported damage found to the unit occurred cannot be determined.The complaint of the coils non-detachment is confirmed even though the coil was returned separated from the dpu, it appears it may have been detached either during handling, cleaning, packaging, or in transit.With the dpu passing electrical testing and without the return of the complete detachment system, the exact root cause of the coils non-detachment cannot be determined, however the evidence as received highly suggests that the most likely contributing factor may have been the incomplete melting of the detachment fiber that adhered to the stretch resistance suture and to coil¿s socket ring.This condition would have prevented the complete detachment of the coil from the dpu.The exact circumstances that produced this condition cannot be determined.In addition, without the return of the complete detachment system and the two microcatheters utilized in the procedure, it cannot be determined if these components contributed to the detachment difficulty.A review of the manufacturing documentation associated with this lot (s10943) presented no issues during the manufacturing or inspection process that can be related to the reported complaint.The complaint of the coils non-detachment is confirmed; the exact root cause of the coils non-detachment cannot be determined, however the evidence as received highly suggests that the most likely contributing factor may have been the incomplete melting of the detachment fiber.Without return of concomitant products it cannot be determined if those components had any additional contributions to the complaint event.Additionally, review of this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.Therefore, no corrective actions will be taken at this time.
|