|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The evaluation of the photos provided from the customer show incorrect orientation of the sphincterotome in the endoscopic view and confirmed the complaint of incorrect orientation.However, without return of the complaint device, further investigation cannot be completed.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Improper cutting wire orientation can occur if the distal end of the catheter is shaped manually.This sphincterotome catheter is precurved and is provided with a precurved stylet in the distal tip of the catheter.This obviates the need for manual formation.The instructions for use contain the following comment: ¿note: do not apply manual pressure to tip or cutting wire of sphincterotome in an attempt to influence orientation, as this may result in damage to device.¿ other factors that can contribute to improper cutting wire orientation include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user: "upon removing device from package, uncoil and straighten sphincterotome."the user is then instructed to "carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." information was provided that prior to use poor orientation was observed at the tip.The instructions for use advise the user: "if package is opened or damaged when received, do not use.Visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use." prior to distribution, all fusion pre-loaded with acrobat wire guide sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.Additional comments regarding this report: based on the information provided that poor orientation was observed prior to use and the device was used, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
|
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion with preloaded acrobat wire guide.When they opened the package and removed the stylet, poor orientation at the tip of the sphincterotome was seen.They tested it [the device] and saw in the papilla the bad orientation [the cutting wire oriented incorrectly to the papilla].
|