(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation determined the reported stent dislodgement and the reported difficulty to remove appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
It was reported that the procedure was to treat a de novo lesion in the non-tortuous, non-calcified, proximal right coronary artery (rca).A 3.50 x 15 mm xience alpine stent delivery system (sds) was selected for the procedure.The sds was advanced to the lesion and crossed.While pulling back the sds to position the stent implant, the edge of the stent implant caught on the end of the guiding catheter and dislodged.An unspecified snare was used in five attempts to retrieve the dislodged stent, however the attempts were unsuccessful.A non-abbott stent implant was used to crush and embed the xience alpine stent implant to the vessel wall.There was a clinically significant delay in the procedure.No additional information was provided.
|