Lot Number 4102411
Medical Device Problem Code Incorrect Or Inadequate Test Results (2456)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2017, the reporter contacted lifescan usa, alleging that the subject meter read inaccurately high compared to another device (true metrics).The reporter claimed obtaining blood glucose readings of ¿367 and 260 mg/dl¿ with the subject meter and ¿214 and 168 mg/dl¿ on the other device, performed within 30 minutes of each other.Based on statistical methodology, the calculated difference of these glucose results exceeds lifescan¿s criteria for accuracy.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO2 METER
Common Device Name
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
gubelstrasse 34
zug 6300
SZ  6300
MDR Report Key6324769
Report Number3008382007-2017-07328
Device Sequence Number13626953
Product Code NBW
UDI-Device Identifier00353885008723
UDI-Public(01)00353885008723
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K131363
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 02/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Lot Number4102411
Other Device ID Number1-2IZSAO8
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date02/03/2017
Device Age21 MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2017
Initial Report FDA Received Date02/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-