| Model Number |
4002-9 |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Device Stops Intermittently (1599)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned sensor was evaluated.During evaluation it was found that the sensor life has expired per patient monitoring time.The sensor passed continuity testing; no open connections or shorts were detected.When connected to a known good philips monitor the sensor displayed a ¿replacesensr" error message.The sensor was determined to be functioning as designed.
|
| |
|
Event or Problem Description
|
|
The customer reported when the sensor is first connected everything works fine for 30-120 minutes; then the reading "goes away." the customer would disconnect and reconnect the sensor to get the reading back, but this strategy no longer worked.It was also indicated that the sensor readings would fluctuate significantly lower to higher and then stabilize.Readings would drop down into the 70's before migrating back up and stabilizing into the mid 90's.No consequences or impact to patient were reported.
|
| |
|