• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC TURBO-JECT® DOUBLE LUMEN POWER-INJECTABLE PICC; LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC TURBO-JECT® DOUBLE LUMEN POWER-INJECTABLE PICC; LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.Reference mdr# 1820334-2017-00288 for associated report.The same patient is involved in both reports.
 
Event or Problem Description
Customer reported that a patient had two previously implanted peripherally inserted central catheter (picc) lines crack and leak at the junction of the hub and catheter within a two week time frame.The picc line was in clinical use for less than one week prior to this event.The device was completely explanted and replaced with a new line.No further patient or event information was provided.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of the complaint history, device history record, instructions for use (ifu), specifications, and quality control was conducted during the investigation.The complaint device was not returned therefore, no physical examinations could be performed; however, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.Review of device history record shows no nonconforming events which could contribute to this failure mode.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBO-JECT® DOUBLE LUMEN POWER-INJECTABLE PICC
Common Device Name
LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key6329466
Report Number1820334-2017-00217
Device Sequence Number10900691
Product Code LJS
UDI-Device Identifier00827002345140
UDI-Public(01)00827002345140(17)180825(10)7227301
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K132334
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberUPICDS-4.0-CT-NT-1111
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/16/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/14/2017
Supplement Report FDA Received Date06/27/2017
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-