| Model Number |
M00558250 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/29/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation findings of distal tip of needle bent.A visual evaluation of the returned device revealed that the syringe luer was cracked.Both the working length of the needle and sheath were kinked at the distal end of the handle and at the distal tip section of the device.Functional evaluation revealed that resistance was met when the needle was extended, furthermore; water was injected into the device and leaking was noted at the syringe luer of the handle.The reported event of luer damage was confirmed.It is most likely handling/manipulation of the device during procedure led to the kinks and bends in the device, therefore the most probable root cause for the complaint event is operational context a review of the device history record (dhr) was performed and no deviations were found.A search of the complaint database confirmed that no similar complaints exist for the specified lot.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a expect pulmonary olympus 25ga needle was used in the airway during an endobronchial ultrasound (ebus) procedure performed on (b)(6) 2016.According to the complainant, during the procedure, a crack was noted at the luer of the handle causing leakage during flushing.A second expect pulmonary olympus device was used to complete the case.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.The investigation found the distal tip of the needle was bent.This is now deemed an mdr reportable event.
|
| |
|
Search Alerts/Recalls
|