MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DORADO PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DORADO PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number DR8094
Medical Device Problem Codes Break (1069); Kinked (1339)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.The lot met all release criteria.Visual/microscopic inspection: the balloon size for this product is printed on the hub and identified the returned sample as a 9mm x 4cm balloon.A complete circumferential shaft break was noted at the proximal butt joint.However, the polyimide appeared to be intact.The edges of the break were examined under microscopic magnification and the edges of the break were slightly jagged.The catheter was kinked at the location of the break, it is likely the user perceived this kink as a bend.Additionally, multiple kinks were noted proximal to the glue joint; however, upon review of the preliminary photos, the manner in which the device was packaged for return may have contributed to the additional kinks.No additional kinks were noted by the user.No other anomalies were observed to the device at this time.Functional/performance evaluation: the patency of the guidewire lumen was tested using an in-house.035" guidewire, and it was able to pass without issues.No further functional testing could be performed due to the break at the proximal balloon glue joint.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: the device was returned.Based on the condition in which the sample was received, the investigation is confirmed for a complete circumferential break at the butt joint and a kinked catheter.Per the reported event details, the device kinked and broke while inserting into a g-tube.Therefore, it is likely procedural issues contributed to the reported event.However, based on available information the definitive root cause is unknown.Labeling review: the current instructions for use (ifu) states: warnings: when the catheter is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended precautions: if resistance is felt during post procedure withdrawal of the catheter through the introducer sheath, determine if contrast is trapped in the balloon with fluoroscopy.If contrast is present, push the balloon out of the sheath and then completely evacuate the contrast before proceeding to withdraw the balloon.If resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and guidewire/introducer sheath as a single unit.Use of the dorado pta dilatation catheter: while maintaining negative pressure and the position of the guidewire, grasp the balloon catheter just outside the sheath and withdraw the deflated dilatation catheter over the wire through the introducer sheath.Use of a gentle clockwise motion may be used to help facilitate catheter removal through the introducer sheath.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that while inserting the pta balloon through a g-tube the balloon allegedly bent and then broke from the catheter shaft.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6330647
Report Number2020394-2017-00068
Device Sequence Number663542
Product Code DQY
UDI-Device Identifier00801741059445
UDI-Public(01)00801741059445(17)191114(10)93YA0067
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/14/2019
Device Catalogue NumberDR8094
Device Lot Number93YA0067
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2017
Initial Report FDA Received Date02/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-