Model Number 1095067
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/23/2016
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this device had low pressure.On receipt of the device for evaluation it was apparent that the mains lead was damaged, leading to exposed wires.The customer was approached with regard to confirming how this damage occurred, but they were unable to provide additional information.The ifu for this device contains the warning 'the power supply cord cannot be replaced by the user.In case the power supply cord becomes damaged, contact philips respironics customer service for replacement (there is no service option)'.No patient harm or death has been reported and the device is not life sustaining/supporting.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSPIRE ELEGANCE
Common Device Name
COMPRESSOR, AIR, PORTABLE
Manufacturer (Section D)
RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
chichester business park
city fields way, tangmere
chichester, west sussex PO20 2FT
UK  PO20 2FT
MDR Report Key6331336
Report Number9681154-2017-00003
Device Sequence Number13657431
Product Code CAF
UDI-Device Identifier00383730001715
UDI-Public00383730001715
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K042655
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1095067
Device Catalogue Number1099973
Device Lot Number140809
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2017
Initial Report FDA Received Date02/15/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-