(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: the device was returned for analysis.Visual and dimensional inspections were performed on the returned device.The stent damage was confirmed.The failure to advance and difficulty removing the device could not be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties were likely due to interactions with the anatomy.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
It was reported that the procedure was to treat a mildly calcified, lesion in the right coronary artery.The 3.5x12 mm xience alpine stent delivery system (sds) failed to cross the target lesion due to the patient anatomy.During removal of the sds, resistance was felt with the anatomy.The xience alpine sds was able to be removed and it was noted that the distal and mid stent struts were flared.The procedure was completed with a same size xience alpine sds.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|