• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. TURON SHOULDER; TURON PRIMARY HUMERAL STEM, SZ 12

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. TURON SHOULDER; TURON PRIMARY HUMERAL STEM, SZ 12 Back to Search Results
Catalog Number 520-01-012
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/23/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the previous surgery failing, hemi shoulder and rotator cuff, the surgeon converted to a reverse.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient had a failed semi shoulder and rotator cuff.The in-vivo length of patient service for the implant was 3.4 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.Initial or prolonged hospitalization was required.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was the result of the patient failed shoulder and rotator cuff.The root cause for the failed shoulder and rotator cuff was not reported.The scope of this investigation is limited without having the explanted parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON PRIMARY HUMERAL STEM, SZ 12
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key6332168
Report Number1644408-2017-00085
Device Sequence Number11142579
Product Code KWS
UDI-Device Identifier00888912024907
UDI-Public(01)00888912024907
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K080402
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Catalogue Number520-01-012
Device Lot Number456G1079
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/23/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/15/2017
Supplement Report FDA Received Date04/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000,LOT 878C1219; 520-46-016,LOT 917C1026
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age77 YR
-
-