| Model Number |
43773 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/12/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Nova biomedical has reached out to the risk manager at the facility to obtain additional information in order to perform a comprehensive investigation.To date, attempts to obtain additional information have not been successful.Unknown if product is available.
|
| |
|
Event or Problem Description
|
|
User-facility medwatch report # (b)(4) received "checked patient glucose prior to dinner and reading was 481.Patient has not been having high glucose.Rechecked the glucose after 1 minute and reading was 280 with same glucometer without any interventions.Glucometer taken out of service.".
|
| |
|
Additional Manufacturer Narrative
|
|
Customer complaint was not confirmed.No product was returned for evaluation.A review was conducted for the lot available for use on or around the time of the event and lot# 0316003309 was identified.A dhr review of the test strip lot was performed.The dhr was complete and contained all relevant data indicating that the released product met all specifications.Retained test strips for this lot were tested as part of the complaint investigation.The test strips met all performance test specification and no discrepant values were obtained.The customers alleged deficiency could not be replicated.
|
| |
|
Search Alerts/Recalls
|