• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS; MONOFOCAL IOLS Back to Search Results
Model Number ZA9003
Medical Device Problem Codes Difficult to Fold, Unfold or Collapse (1254); Kinked (1339)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age/date of birth: unknown/ not provided.Sex/gender: unknown/ not provided.Date of event: unknown/not provided.If implanted; give date: n/a (not applicable).The intraocular lens was removed and replaced during the same procedure.If explanted; give date: n/a (not applicable).The intraocular lens was removed and replaced during the same procedure.Device evaluation: the product was not returned; therefore, the reported issue was not verified manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provide instructions and precautions along with warnings for the proper use and handling of the device.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
According to the surgeon, the intraocular lens (iol) had a crease in its center when it unfolded.The surgeon loaded the iol and began to insert it into the eye.As the lens was unfolding inside the eye, he could see that it didn't look right under the microscope.It was stated that the iol had a crease in it and was not unfolding correctly.The iol was removed and replaced with a different one during the same procedure.The incision did not have to be enlarged.The patient was not injured.The patient is doing fine.No additional information was provided to abbott medical optics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6333267
Report Number2648035-2017-00303
Device Sequence Number1016541
Product Code HQL
UDI-Device Identifier05050474528925
UDI-Public(01)05050474528925(17)180801
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/01/2018
Device Model NumberZA9003
Device Catalogue NumberZA90030155
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2017
Initial Report FDA Received Date02/15/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EMERALDC30, LOT CB39363
-
-