| Catalog Number |
PC410051730 |
| Medical Device Problem Codes |
Difficult or Delayed Positioning (1157); Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/24/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The presidio will not be returned, therefore the root cause of the positioning and detachment difficulty cannot be determined.Dhr: a review of the manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.Based on the information, the event could not be confirmed.The product was not returned for analysis; however a review of the manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.Therefore, no corrective actions will be taken at this time.
|
| |
|
Event or Problem Description
|
|
The contact from the facility reported that during coil embolization, the presidio coil (pc410051730/c38467) detached after pressing the detach button several times on the enpower dcb.However this resulted in an inaccurate placement of the last loop of the coil as the user moved the microcatheter several times to check if the coil was detached.The last loop of the coil was therefore outside the aneurysm in the parent vessel.The physician tried to push it back into the aneurysm with the microcatheter but was unsuccessful.No further intervention was taken with respect to the coil protrusion.There was no flow restriction as a result.The other coils were detached properly.There was not a significant delay to the procedure as a result of the issue.The coil detached in a few minutes.There were no damages noted to the coil/coil delivery system or detachment system.A pre-deployment electrical check was performed and there was no fault light and the green system ready light illuminated.During the detachment cycle the detachment light illuminated.There was an audible beep at the time of the detachment attempt.All connections appeared to fit properly without use of excessive force.
|
| |
|
Search Alerts/Recalls
|