• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY STRATA NSC LP SHUNT KIT; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY STRATA NSC LP SHUNT KIT; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Catalog Number 44420
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/16/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Only the lumbar catheter was returned.The catheter was patent.However, it did not meet the requirements for leak testing.The catheter was observed to be sheared off the distal end.It is unknown how or when the damage occurred.The catheter met the requirements for the tensile strength and elongation tests.The instructions for use that accompany the device cautions that low tear strength is a characteristic of most unreinforced silicone elastomer materials, and that care must be taken with the handling and placement of the silicone elastomer catheter tubing to avoid cuts, nicks, or tears.It also cautions that to avoid possible transection of the catheter during catheter placement, the catheter should never be withdrawn through the tuohy needle.If the catheter needs to be withdrawn, the tuohy needle, guidewire (if used), and the catheter must be removed simultaneously.There was proteinaceous debris observed within the interior and exterior of the catheter.A review of the manufacturing records showed no anomalies.All catheters are 100% inspected at the time of manufacture.
 
Event or Problem Description
It was reported the catheter ruptured intra-operatively.According to the report, another catheter was used to complete the procedure and there was no injury to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA NSC LP SHUNT KIT
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key6335819
Report Number2021898-2017-00058
Device Sequence Number16399429
Product Code JXG
UDI-Device Identifier00643169010796
UDI-Public00643169010796
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K033850
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Catalogue Number44420
Device Lot NumberE18455
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2017
Initial Report FDA Received Date02/16/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-