| Catalog Number |
C45101796D-NL |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Customer has not yet returned the device to the manufacturer for device evaluation.When and if the device becomes available and is returned and evaluated the manufacturer will file a follow-up report detailing the results of the evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the during use of a disposable portex® anesthesia breathing circuit, the co2 port broke while turning the heat and moisture exchanger (hme).According to the reporter, the event occurred in the operating room while the device was connected to the patient.The hme was replaced.No patient injury occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
No product was returned.Instead a picture was sent.From the picture, it can be observed a portex 120-1500ml foam hme from pn 9064/711 with the gas monitoring port broke.The customer's reported problem was "the or was using one of the disposable anesthesia breathing circuits and the co2 port broke while turning the hme".No testing could be performed because no samples were provided for evaluation.A review of the manufacturing and quality inspection processes were reviewed and considered adequate and correct.The packaging operations were reviewed and considered adequate and correct.During verification, 32 assemblies were reviewed in order to verify that hme assemblies are free of damage such as broken parts.None of the 32 samples had any defects.The most probable root cause was inadequate visual inspection to detect damaged hme assembly.
|
| |
|