| Model Number |
MT20649-PNK |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/24/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(6) 2017 that on (b)(6) 2017 the receiver had no audio output.No additional event or patient information is available.The receiver was returned for evaluation.An exterior visual inspection was performed and passed.A manual test was performed and failed.The reported event of no audio output was confirmed.The root cause was determined to be a defective speaker.
|
| |
|
Search Alerts/Recalls
|