Product analysis (pa) for the returned generator was completed.The reported ¿extrusion/generator¿ allegation is beyond the scope of the activities performed in the pa laboratory.During analysis, it was found that the device performed according to functional specifications and concluded that no abnormal performance, or any other types of adverse conditions, was found.It was later reported by the physician's office that they believe the patient had picked at the incision, which caused the extrusion.Pa for the returned lead was completed.The allegation of extrusion is not an actionable issue which will result in confirmation of an event and identification of a casual factor in the pa laboratory environment.Note that the lead assembly, including the electrode section, was not returned for analysis and thus a complete evaluation could not be performed on the entire lead product.The condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.Based on the findings in the product analysis lab, there is no evidence to suggest an anomaly with the returned portions of the device.
|