Information was received from a manufacturer representative (rep) regarding a patient who was implanted with a neurostimulator (ins) for post laminectomy pain.It was reported that impedance issues were noted with the patient¿s ins, and the patient was due for an ins replacement.Impedance measurements were ran at 3.0v and electrode pairs c<(>&<)>1, c <(>&<)>2, 0<(>&<)>1, 0<(>&<)>2, 1<(>&<)>2, 1<(>&<)>3, and 2<(>&<)>3 were all >4,000 ohms.C<(>&<)>0 and 0<(>&<)>3 were 874 ohms, while the measurement for c<(>&<)>3 gave an invalid ¿???¿ value.No symptoms were reported.Additional information from the rep reported that the cause of the high impedance was not determined.It was noted that the troubleshooting performed was that they tried increasing the voltage.A surgery was scheduled to replace the ins to resolve the issue.Due to the patient¿s age and adequate coverage, the current paddle lead was going to be left in place.
|