• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES SPINE CONFIDENCE KIT SPINAL CEMENT SYSTEM - NO NEEDLE; CONFIDENCE CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES SPINE CONFIDENCE KIT SPINAL CEMENT SYSTEM - NO NEEDLE; CONFIDENCE CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Catalog Number 283913000
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/25/2017
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).A follow up report will be filed upon completion of the investigation.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The cement delivery hydraulic device exploded twice.One (b)(6) and again to my notification on (b)(6).
 
Manufacturer Narrative
One (1) confidence kit no needles [product code: 2839-13-000] was returned to the customer quality unit (cqu).For evaluation.Visual examination of the complaint sample revealed that the confidence pump body cracked at the distal tip, where the flex tube exits the part.The crack initiated at the tip and propagated to a length of approximately 2.5 inches.Additionally, it should be noted that the crack formed through the middle of the pump body approximately 70-80 degrees from the parting line.At manufacturing the pump undergoes a 100% high pressure testing in-process where the safety relief valve is actuated ensuring proper functionality and prevention of pump cracking.A definitive root cause cannot be determined at this time.A potential root cause for this failure is that the safety relief valve did not actuate properly to relieve pressure, and that the pump body inner diameter may have been too small.Without the lot number manufacturing record review could not be completed.No systemic trends requiring immediate action have been observed, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIDENCE KIT SPINAL CEMENT SYSTEM - NO NEEDLE
Type of Device
CONFIDENCE CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
jason busch
325 paramount drive
raynham, MA 02767
5088808201
MDR Report Key6347240
MDR Text Key67934814
Report Number1526439-2017-10133
Device Sequence Number1
Product Code NDN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number283913000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/25/2017
Initial Date FDA Received02/21/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/17/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-