• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. OMEGA II HIP; NECK, FEMORAL, MODULAR, 32MM 8 DEGREE CERASIVE TAPER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. OMEGA II HIP; NECK, FEMORAL, MODULAR, 32MM 8 DEGREE CERASIVE TAPER Back to Search Results
Catalog Number 410-32-108
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/30/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the r120 modular neck disconnecting and causing the hip to dislocate.A neck replacement was not an option for the r120, so a full revision was completed with a lima.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was the patient's r120 modular neck disconnected and caused the hip to dislocate.The in-vivo length of patient service for the implant was 3.6 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.Initial or prolonged hospitalization was required.The revision surgery was completed as intended.The device was reported to have been kept by the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was the result of the disassociation of the r120 modular neck causing the hip to dislocate.The root cause for the disassociation was not reported.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMEGA II HIP
Common Device Name
NECK, FEMORAL, MODULAR, 32MM 8 DEGREE CERASIVE TAPER
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6347242
Report Number1644408-2017-00134
Device Sequence Number16384833
Product Code LPH
UDI-Device Identifier00888912085861
UDI-Public(01)00888912085861
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K000817
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2019
Device Catalogue Number410-32-108
Device Lot Number896B1044
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/30/2017
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date02/21/2017
Supplement Report FDA Received Date04/14/2017
06/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
411-00-000, LOT 902B1097
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age53 YR
-
-