| Model Number |
2261 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
12/30/2016
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Instrumentation laboratory (il) conducted an investigation that included a review of data files from the customer's gem premier 4000.The data review showed that the device was operating within the specifications.There was no system or sensor malfunction during analysis of the patient's samples.The customer stated that there was no correlation of this patient's death to gem premier 4000 instrument performance.Therefore, no remedial action is required at this time.
|
| |
|
Event or Problem Description
|
|
A customer requested instrumentation laboratory (il) to review two glucose recoveries from a gem premier 4000 instrument.Both samples were from the same patient.The samples were analyzed on (b)(6) 2016.The first sample analyzed at 12:56 produced an incalculable glucose result.The second sample analyzed at 13:37 produced a glucose result of >41.6 mmol/l.Il was informed that the patient died later.
|
| |
|