| Catalog Number |
9731965 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/25/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An onsite medtronic representative was onsite when this event occurred, they confirmed that the instrument was damaged.Replacement of the instrument resolved the issue.No further issues were reported.Analysis of the returned instrument confirmed the physical damage as the issue was replicated due to a bent support arm yielding high geometry error.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that while in a cranial resection procedure, one of the spheres on the probe was showing red in the camera tracking view of the navigation system.They switched spheres and it still showed red.When examining the probe closer it appeared that a post was bent.They attempted to bend it back, but afterward the probe would no longer verify.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no impact on patient outcome.
|
| |
|