Catalog Number C-HSK-3038 |
Device Problem
Positioning Failure (1158)
|
Patient Problem
No Information (3190)
|
Event Date 01/24/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).A lot history record review was completed for the reported product lot number.There was no nonconformance recorded in the lot history the device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
|
|
Event Description
|
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system 3.8 mm did not deploy.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
|
|
Manufacturer Narrative
|
(b)(4).The device was returned to the factory for evaluation.Signs of clinical use and no evidence of blood were observed.The delivery device was received inside the loading device.There was no blood in or on the delivery tube.The slide lock was engaged and the plunger was not depressed on the delivery device.The seal and tension spring assembly remained inside the loading device.The following measurements of the delivery tube were taken : the inner delivery tube diameter was measured at.198 inches , the outer diameter was measured at.220 inches.The length of the delivery tube was measured at 2.50 inches.The values recorded were within the tolerance specifications.The seal was removed from the loading device.There were no cracks or delamination's observed.The seal was wrapped (folded) indicating that the seal was rolled correctly.Based upon the received condition of the device, the reported complaint was not confirmed for the reported failure mode "failure to deploy" but was confirmed for the analyzed failure mode "fitting problem".
|
|
Event Description
|
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system 3.8 mm did not deploy.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
|
|