• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CV HS III PROXIMAL SEAL SYTEM 3.8MM; CLAMP, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CV HS III PROXIMAL SEAL SYTEM 3.8MM; CLAMP, VASCULAR Back to Search Results
Catalog Number C-HSK-3038
Device Problem Positioning Failure (1158)
Patient Problem No Information (3190)
Event Date 01/24/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A lot history record review was completed for the reported product lot number.There was no nonconformance recorded in the lot history the device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system 3.8 mm did not deploy.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
 
Manufacturer Narrative
(b)(4).The device was returned to the factory for evaluation.Signs of clinical use and no evidence of blood were observed.The delivery device was received inside the loading device.There was no blood in or on the delivery tube.The slide lock was engaged and the plunger was not depressed on the delivery device.The seal and tension spring assembly remained inside the loading device.The following measurements of the delivery tube were taken : the inner delivery tube diameter was measured at.198 inches , the outer diameter was measured at.220 inches.The length of the delivery tube was measured at 2.50 inches.The values recorded were within the tolerance specifications.The seal was removed from the loading device.There were no cracks or delamination's observed.The seal was wrapped (folded) indicating that the seal was rolled correctly.Based upon the received condition of the device, the reported complaint was not confirmed for the reported failure mode "failure to deploy" but was confirmed for the analyzed failure mode "fitting problem".
 
Event Description
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system 3.8 mm did not deploy.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HS III PROXIMAL SEAL SYTEM 3.8MM
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CV
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CV
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
45 barbour pond drive
wayne, NJ 07470
MDR Report Key6347719
MDR Text Key68219104
Report Number2242352-2017-00171
Device Sequence Number1
Product Code DXC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/08/2017
Device Catalogue NumberC-HSK-3038
Device Lot Number25129394
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/24/2017
Initial Date FDA Received02/21/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-