| Model Number |
ZM-540PA |
| Medical Device Problem Codes |
Device Alarm System (1012); No Audible Alarm (1019); Alarm Not Visible (1022)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/27/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The biomedical engineer reported that the transmitter did not v-fib alarm while the patient was being monitored on the cns.The device was in use on a patient, however no patient harm was reported.The customer sent the device in and was provided with a transmitter exchange.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer reported that the transmitter did not v-fib alarm while the patient was being monitored on the cns.
|
| |
|
Additional Manufacturer Narrative
|
|
The biomedical engineer reported that the transmitter did not v-fib alarm while the patient was being monitored on the cns.The device was in use on a patient; however no patient harm was reported.The customer sent the device in for evaluation and was provided with an exchanged device.The unit was cleaned and evaluated and the reported problem was not duplicated.The unit completed 24 hours of extended testing and operates to manufacturer's specifications.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
|
| |
|