• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TURON SHOULDER; TURON PRIMARY HUMERAL STEM, SZ 6

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TURON SHOULDER; TURON PRIMARY HUMERAL STEM, SZ 6 Back to Search Results
Catalog Number 520-01-006
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/31/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Third revision surgery - due to the patient having a chronic infection.
 
Additional Manufacturer Narrative
The reason for this revision surgery was a chronic infection.The previous surgery and the revision detailed in this investigation occurred 21 days apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.The healthcare professional indicated there was a serious risk to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post surgical instructions.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON PRIMARY HUMERAL STEM, SZ 6
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6350742
Report Number1644408-2017-00095
Device Sequence Number11062743
Product Code KWS
UDI-Device Identifier00888912144896
UDI-Public(01)00888912144896
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K080402
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/28/2019
Device Catalogue Number520-01-006
Device Lot Number533G1028
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/31/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/22/2017
Supplement Report FDA Received Date03/28/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000, LOT 878C1637; 520-46-024, LOT 921C1016
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age57 YR
Patient Weight88
-
-