MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number U35130312
Medical Device Problem Codes Break (1069); Inflation Problem (1310); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/22/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly failed to inflate during the first inflation.Reportedly, the balloon catheter was exchanged over the guide wire for another balloon and the procedure was completed.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Visual/microscopic inspection: the balloon appeared to have been previously inflated.The balloon size for this product is printed on the balloon hub of the catheter and identified the returned sample as a 3 mm x 120 mm balloon.A complete circumferential catheter detachment was identified at the bifurcate.No other anomalies were noted to the strain relief or the y-hub.Functional/performance evaluation: the strain relief was removed from the y-hub.The catheter detachment was examined under microscopic magnification (10x) and the edges of the break were jagged.Sanding marks were noted on the catheter at the location of the break.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: the device was returned.The investigation is confirmed for a complete circumferential catheter detachment just distal to the y-hub.The investigation is inconclusive for inflation issues, as functional testing could not be performed due to the poor sample condition.Excessive sanding of the catheter under the strain relief is the root cause for the catheter detachment.It is likely that the leaking from the detachment caused the reported inflation issues.Corrective and preventative actions have been implemented, this lot is pre corrective action.Labeling review: the current ifu (instructions for use) states: warnings: - when the catheter is exposed to the vascular system, it should be manipulated while under high quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip or catheter breakage, catheter kink, or balloon separation.Precautions: - if resistance is felt during post procedure withdrawal of the catheter through the introducer sheath/guide catheter, determine if contrast is trapped in the balloon with fluoroscopy.If contrast is present, push the balloon out of the introducer sheath/guide catheter and then completely evacuate the contrast before proceeding to withdraw the balloon.- if resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and introducer sheath/guide catheter as a single unit.Potential adverse reactions the complications that may result from a peripheral balloon dilatation procedure include: - additional intervention use of ultraverse 035 pta dilatation catheter: - apply negative pressure to fully evacuate fluid from the balloon.Confirm that the balloon is fully deflated under fluoroscopy and that no contrast is left in the balloon.- while maintaining negative pressure and the position of the guidewire, grasp the balloon catheter just outside the introducer sheath/guide catheter and withdraw the deflated dilatation catheter over the wire through the introducer sheath/guide catheter.Use of a gentle counterclockwise motion may be used to help facilitate catheter removal through the introducer sheath/guide catheter.(b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly failed to inflate during the first inflation.Reportedly, the balloon catheter was exchanged over the guide wire for another balloon and the procedure was completed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035 PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6351750
Report Number2020394-2017-00100
Device Sequence Number635774
Product Code LIT
UDI-Device Identifier00801741091711
UDI-Public(01)00801741091711(17)171231(10)50118783
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2016
Device Catalogue NumberU35130312
Device Lot Number50118783
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 01/26/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/22/2017
Supplement Report FDA Received Date03/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
Patient Weight72
-
-