| Model Number |
ZXT150 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Flashers (1864); Vitreous Floaters (1866); Blurred Vision (2137)
|
| Date of Event |
01/24/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the intraocular lens (iol) was explanted as patient needed a different power lens due to flashes/floaters and blurry vision.The lens was replaced with a smaller, 15.5 diopter lens.There was no incision enlargement, vitrectomy, and no sutures required in the explant surgery.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes.Returned to manufacturer on: 03/08/2017.Device returned to manufacturer? yes.The intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection showed that the lens was cut in half, most probably to make the explant possible.Considering the condition of the lens additional analysis was not possible.The customer's reported complaint could not be verified.The manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.The directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Search Alerts/Recalls
|