| Catalog Number |
48922601 |
| Medical Device Problem Codes |
Material Deformation (2976); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/27/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; one of the blades bent during insertion.
|
| |
|
Additional Manufacturer Narrative
|
|
Device history review; complaint history review; risk assessment; manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Visual, dimensional and functional analysis could not be performed as the device was not returned.The most likely cause of the reported event was determined to be related tol pilot cutter was not used prior to the anchor insertion, which was recommended per surgical technique.
|
| |
|
Event or Problem Description
|
|
It was reported that; one of the blades bent during insertion.
|
| |
|