ACE SURGICAL SUPPLY COMPANY EXTERNAL HEX IMPLANT; DENTAL IMPLANT Back to Search Results
Model Number 403713
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/03/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event has been determined to be a post-placement/pre-load implant failure.This implant loss suggests that the source of the problem was likely to stem from procedural errors and/or the lack of osseointegration.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, or patient behavior.Proper intended use of the implant is described in its instructions for use.Physical analysis was not performed.
 
Event or Problem Description
Mobility, inadequate bone quality and pain were reported.Time of loss in relation to the implantation was before prosthetic restoration.Primary stability and osseointegration was achieved.Bone quality at the time of implant failure type iv.Patient was undergoing chemotherapy at the time of implantation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTERNAL HEX IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ACE SURGICAL SUPPLY COMPANY
1034 pearl street
brockton MA 02301
MDR Report Key6353333
Report Number1287163-2017-50851
Device Sequence Number1041850
Product Code DZE
UDI-Device Identifier00614950005437
UDI-Public00614950005437
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K954513
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 02/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dentist
Device Model Number403713
Device Catalogue Number403713
Device Lot Number14050080
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2017
Initial Report FDA Received Date02/23/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-