| Model Number |
A2-80072 |
| Medical Device Problem Codes |
Split (2537); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
01/25/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer has received the affected iv set on 02/17/2017.The returned iv set is being sent to the contract supplier for evaluation.The manufacturer will actively follow up on this investigation.
|
| |
|
Event or Problem Description
|
|
The user reported that "we went to infuse a patient with tysabri on one of our zyno pumps.The pump set had a split in the actual tubing and the medication spilled out on the floor of the clean room." on the follow-up call with a zyno customer service employee, the customer mentioned that the iv set was leaking in the mixing room and was never placed in the pump.No patient injury or harm occurred for this incident.
|
| |
|
Additional Manufacturer Narrative
|
|
The user reported issue was confirmed.The iv set was evaluated by zyno medical and was found to have a split-tubing due to manufacturing defect.Details of the testing can be found in the report.A quality alert has been sent to the contract manufacturer on 04/13/2017.
|
| |
|
Event or Problem Description
|
|
This is a follow-up for the initially filed mdr (3006575795-2017-00051).
|
| |
|