• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO JOEY,PUMP W/POLE CLMP; ENTERAL FEEDING PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO JOEY,PUMP W/POLE CLMP; ENTERAL FEEDING PUMP Back to Search Results
Model Number 383400
Device Problems Crack (1135); No Display/Image (1183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 02/17/2017.An investigation is currently under way.Upon completion the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2017 that a customer had an issue with a feeding pump.The customer reports the unit has lcd issues.On (b)(6) service tech confirmed complaint.Upon further information from service tech as for this unit like most cracked lcds the display was black along the cracks and had very minimal sections of the screen that were legible i would say about 75-80% was blank.Once these liquid crystal displays break they don't really leave anything to be seen and because of the small size they are almost always blank.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed for the reported condition of lcd issues.The unit was triaged and the reported issue was confirmed.The pump was excessively damaged, so the root cause is categorized as customer misuse.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEY,PUMP W/POLE CLMP
Type of Device
ENTERAL FEEDING PUMP
Manufacturer (Section D)
COVIDIEN
16 tampines industrial crecsent
singapore
SN 
Manufacturer (Section G)
COVIDIEN
16 tampines industrial crecsent
singapore
SN  
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 04048
5084524151
MDR Report Key6360169
MDR Text Key68608734
Report Number1282497-2017-00230
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number383400
Device Catalogue Number383400
Device Lot NumberF14051043
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/13/2016
Initial Date FDA Received02/27/2017
Supplement Dates Manufacturer Received12/13/2016
Supplement Dates FDA Received10/26/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-