| Model Number |
D-1336-03-S |
| Health Effect - Clinical Codes |
Low Blood Pressure/ Hypotension (1914); Cardiac Tamponade (2226)
|
| Date of Event |
02/09/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the product was not returned for analysis, no product failure analysis can be conducted and no determination of possible contributing factors could be made.As such, the investigation will be closed.If the complaint device is received in the future we will reopen the complaint and perform the investigation as appropriate.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an ablation procedure for atrial fibrillation with a thermocool smarttouch uni-directional navigation catheter and suffered a cardiac tamponade requiring pericardiocentesis.During the procedure, the patient became hypotensive and a tamponade was confirmed via ultrasound.Pericardiocentesis was in progress at the time of complaint report.There is no information regarding extended hospitalization or patient outcome.Patient was reported to be in stable condition.Physician did not provide a causality opinion.Multiple attempts have been made to obtain clarification to this complaint.However, no further information has been made available.
|
| |
|
Additional Manufacturer Narrative
|
|
The bwi failure analysis lab received the device for evaluation on 03/03/2017.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).It was reported that a patient underwent an ablation procedure for atrial fibrillation with a thermocool® smarttouch® uni-directional navigation catheter and suffered a cardiac tamponade requiring pericardiocentesis.The returned device was visually inspected and it was found in normal conditions.Per the event, the catheter was tested for electrical performance, temperature response and generator test and it was found within specifications.A deflection test was performed and the catheter passed.The catheter was evaluated for eeprom, and the functionality of the sensors of the catheter were tested on carto3 system.The catheter was recognized by carto 3 system, no error messages were displayed and the catheter was properly visualized.Eeprom data demonstrates the catheter was properly calibrated during manufacturing.The force features was evaluated and pass.Finally, an irrigation test was performed and the catheter passed, no occlusion was observed.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.The catheter passed all specifications.The root cause of the cardiac tamponade remains unknown.The instructions for use states that careful catheter manipulation must be performed in order to avoid cardiac damage, perforation or tamponade.
|
| |
|