• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENTEROMEDICS, INC. MAESTRO RECHARGEABLE SYSTEM; RECHARGEABLE NEUROREGULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENTEROMEDICS, INC. MAESTRO RECHARGEABLE SYSTEM; RECHARGEABLE NEUROREGULATOR Back to Search Results
Model Number 2002
Medical Device Problem Codes Device Inoperable (1663); Use of Device Problem (1670); Low impedance (2285); Device Displays Incorrect Message (2591); Impedance Problem (2950)
Health Effect - Clinical Codes Failure of Implant (1924); Twiddlers Syndrome (2114); Therapeutic Response, Decreased (2271)
Date of Event 09/21/2016
Type of Reportable Event Serious Injury
Event or Problem Description
A former subject in the empower clinical trial ((b)(4); subject id (b)(6)) was originally converted to a maestro rechargeable system on (b)(6) 2015.The system, including model 2002 rechargeable neuroregulator, model 2200a-47e anterior lead, and model 2200p-47e posterior lead were replaced on (b)(6) 2016 (refer to mdr3005025697-2017-00005 through mdr3005025697-2017-00007 for details on original rechargeable system removal due to excessive manipulation).On (b)(6) 2016, the site reported that alerts for low impedance were experienced at follow-up, noting that the ring electrodes were placed quite close together at implant.On (b)(6) 2016, it was reported that low impedance measurements were again experienced.Radiograph review suggested that the ring electrodes were in very close proximity to each other with twisting of the leads near the neuroregulator.The replacement system was explanted on (b)(6) 2016.Mdrs related to this issue are: mdr3005025697-2017-00008: model 2002 rechargeable neuroregulator.Mdr3005025697-2017-00009: model 2200a-47e anterior lead.Mdr3005025697-2017-00010: model 2200p-47e posterior lead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAESTRO RECHARGEABLE SYSTEM
Common Device Name
RECHARGEABLE NEUROREGULATOR
Manufacturer (Section D)
ENTEROMEDICS, INC.
2800 patton road
saint paul MN 55113
MDR Report Key6370106
Report Number3005025697-2017-00008
Device Sequence Number19122637
Product Code PIM
UDI-Device Identifier00857334004262
UDI-Public00857334004262
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P130019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date12/01/2016
Device Model Number2002
Device Catalogue Number2002
Device Lot Number093513215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2017
Initial Report FDA Received Date03/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BUSCOPAN; IBUPROFEN; OXYCODONE; PARACETAMOL
Outcome Attributed to Adverse Event Required Intervention;
Patient Age42 YR
Patient Weight122
-
-