TORAX MEDICAL, INC. LINX REFLUX MANAGEMENT SYSTEM; ANTI-REFLUX IMPLANT Back to Search Results
Model Number LS15
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Heartburn/Indigestion (1883)
Date of Event 02/10/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Following a laparoscopic anti-reflux procedure, a patient experienced ongoing gerd symptoms resulting in explant of the linx device.The linx device was used as part of the anti-reflux procedure.Anti-reflux procedure and linx device implantation occurred without issue (b)(6) 2014.Patient was symptom free until the return of gerd symptoms 4 months prior to removal.Patient had an upper gi on (b)(6) 2017 and a linx bead was believed to have eroded through the patient's esophagus.Device explant due to ongoing gerd on (b)(6) 2017; the device had not eroded through the esophagus.A foreign body (including blue monofilament suture) unassociated with the linx device was found approximately 1 cm from the gastroesophageal junction.The device was found in the correct position/geometry at the time of removal.Patient underwent a fundoplication immediately after device explant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINX REFLUX MANAGEMENT SYSTEM
Common Device Name
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4188 lexington avenue n
shoreview MN 55126
MDR Report Key6370418
Report Number3008766073-2017-00027
Device Sequence Number443511
Product Code LEI
UDI-Device Identifier00855106005042
UDI-Public00855106005042
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P100049
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date07/01/2017
Device Model NumberLS15
Device Lot Number4925
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2017
Initial Report FDA Received Date03/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient Age58 YR
-
-