• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CFF03, 5X100 KII FIOS ADVFIX 6/BX; GCJ

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CFF03, 5X100 KII FIOS ADVFIX 6/BX; GCJ Back to Search Results
Model Number 24
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/08/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event product was not returned to applied medical for evaluation.The root cause of the event is likely due excessive force applied during insertion.Applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.This report is being filed as a result of a re-review of applied medical complaints received between june 1, 2014 and may 31, 2016.This retrospective review was associated with a quality management system (qms) compliance action plan developed and presented to fda to address an april 10, 2015 warning letter.Applied medical has revised its mdr reporting criteria to be more conservative and has improved complaint handling and mdr reporting processes.The reviews ensured that recent reportable events were appropriately identified and reported to the designated regulatory authority(ies).This report, which represents the initial and final reports combined, is being submitted based on the findings of that retrospective review.In accordance with 21 cfr 803.56, if additional information is obtained which was not known or was not available when this report was submitted, then a supplemental report will be submitted to the fda.
 
Event or Problem Description
"dr.(b)(6) was inserting trocar into patient and trocar broke.Inner obturator broke in half and cannula bent.Hospital has trocar in sealed bag to send back to applied." patient status - no impacted by incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CFF03, 5X100 KII FIOS ADVFIX 6/BX
Common Device Name
GCJ
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key6370845
Report Number2027111-2017-01183
Device Sequence Number16408103
Product Code GCJ
UDI-Device Identifier00607915117641
UDI-Public(01)00607915117641(17)180624(30)01(10)1249147
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
24
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/24/2018
Device Model Number24
Device Catalogue Number24
Device Lot Number1249147
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2015
Initial Report FDA Received Date03/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-