U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CFF71, 12X150 KII FIOS ADVFIX 6/BX; GCJ

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CFF71, 12X150 KII FIOS ADVFIX 6/BX; GCJ Back to Search Results
Model Number 5
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/16/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event product was returned to applied medical for evaluation with 1 rubber fragment.Upon inspection, engineering noted fragmentation of an internal rubber component of the seal.The rubber fragment returned matched the missing portion on the seal.The likely root cause of the damage to the seal is an instrument.Applied medical's instructions for use states that "extra care should be used when inserting angular and asymmetrical instruments, such as 'j' hooks and clip appliers.All instruments should be centered axially when inserted through the seal to prevent tearing." applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.This report is being filed as a result of a re-review of applied medical complaints received between june 1, 2014 and may 31, 2016.This retrospective review was associated with a quality management system (qms) compliance action plan developed and presented to fda to address an april 10, 2015 warning letter.Applied medical has revised its mdr reporting criteria to be more conservative and has improved complaint handling and mdr reporting processes.The reviews ensured that recent reportable events were appropriately identified and reported to the designated regulatory authority(ies).This report, which represents the initial and final reports combined, is being submitted based on the findings of that retrospective review.In accordance with 21 cfr 803.56, if additional information is obtained which was not known or was not available when this report was submitted, then a supplemental report will be submitted to the fda.
 
Event or Problem Description
Robotic laparoscopic hysterectomy - "case was performed by dr.(b)(6).A cff73 trocar was used for her camera port.During the case a small piece of black plastic was observed inside of the patient's abdomen.It was removed from the patient and appeared to be a piece of the trocar seal.Visual inspection shows that the piece appears to have been torn from the seal, rather than cut or melted, because it has a rough edge.The only instrument that has passed through the trocar was the robotic camera." patient status - at home with no issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CFF71, 12X150 KII FIOS ADVFIX 6/BX
Common Device Name
GCJ
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key6370901
Report Number2027111-2017-01235
Device Sequence Number19024701
Product Code GCJ
UDI-Device Identifier00607915123048
UDI-Public(01)00607915123048(17)180608(30)01(10)1248320
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
5
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/08/2018
Device Model Number5
Device Catalogue Number5
Device Lot Number1248320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2015
Initial Report FDA Received Date03/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-