• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CTB03, 5X100 KII SH BLD ZTHR 6/BX; GCJ

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CTB03, 5X100 KII SH BLD ZTHR 6/BX; GCJ Back to Search Results
Model Number 3
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/25/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event unit was returned for evaluation.Engineering was able to replicate the customer's experience and discovered internal damage.The root cause was the internal damage.Engineering was able to correct the damage and the device functioned properly.Internal damage was likely caused by improper use.Applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.This report is being filed as a result of a re-review of applied medical complaints received between june 1, 2014 and may 31, 2016.This retrospective review was associated with a quality management system (qms) compliance action plan developed and presented to fda to address an april 10, 2015 warning letter.Applied medical has revised its mdr reporting criteria to be more conservative and has improved complaint handling and mdr reporting processes.The reviews ensured that recent reportable events were appropriately identified and reported to the designated regulatory authority(ies).This report, which represents the initial and final reports combined, is being submitted based on the findings of that retrospective review.In accordance with 21 cfr 803.56, if additional information is obtained which was not known or was not available when this report was submitted, then a supplemental report will be submitted to the fda.
 
Event or Problem Description
Procedure performed unknown."one two separate intra-operative occasion the sliding lock (safety shield) on the ctb03 trocar would not engage.Slide slided back to neutral position.Two devices detected so far." patient status: no adverse events reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CTB03, 5X100 KII SH BLD ZTHR 6/BX
Common Device Name
GCJ
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key6372195
Report Number2027111-2017-01217
Device Sequence Number932285
Product Code GCJ
UDI-Device Identifier00607915123512
UDI-Public(01)00607915123512(17)180623(3))01(10)1249103
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
3
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/23/2018
Device Model Number3
Device Catalogue Number3
Device Lot Number1249103
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2015
Initial Report FDA Received Date03/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-