• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493808012320
Medical Device Problem Code Kinked (1339)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/28/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at time of event: 18 years or older (b)(4).Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The balloon was loosely folded.There was contrast in the inflation lumen.There was blood in between balloon folds.The outer shaft, inner shaft, balloon and tip were microscopically examined.The hypotube shaft was completely separated 56.5 cm from the tip.The fracture faces were oval as if kinked prior to separation.There were numerous hypotube kinks.There was tip damage.There was no evidence of any material or manufacturing deficiencies contributing to the damage.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event or Problem Description
Reportable based on device analysis completed on 08-feb-2017.It was reported that shaft kink occurred.The target lesion was located in a coronary artery.A 3.25mm x 12mm quantum¿ maverick¿ balloon catheter was advanced for dilation.However, it was noted that the shaft was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed hypotube break.It was further reported that the shaft broke during advancing of the device to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM¿ MAVERICK¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key6372636
Report Number2134265-2017-01340
Device Sequence Number943465
Product Code LOX
UDI-Device Identifier08714729392682
UDI-Public(01)08714729392682(17)20190630(10)19342312
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2019
Device Model NumberH7493808012320
Device Catalogue Number38080-1232
Device Lot Number19342312
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2017
Initial Report FDA Received Date03/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-