Model Number 2
Medical Device Problem Code Physical Resistance (2578)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/10/2014
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: the event product was returned to applied medical for evaluation.Upon inspection, engineering noticed the septum was stretched.Engineering attempted to replicate the customer's experience of drag, but could not.The root cause of the septum stretching is due to the separator being inserted into the trocar system for an undefined time period.Even though the root cause of the drag could not be determined, applied medical continuously seeks to improve the form, function and ease of use of its products and will continue to monitor its vigilance system for trends and take appropriate actions as necessary.This report is being filed as a result of a re-review of applied medical complaints received between june 1, 2014 and may 31, 2016.This retrospective review was associated with a quality management system (qms) compliance action plan developed and presented to fda to address an april 10, 2015 warning letter.Applied medical has revised its mdr reporting criteria to be more conservative and has improved complaint handling and mdr reporting processes.The reviews ensured that recent reportable events were appropriately identified and reported to the designated regulatory authority(ies).This report, which represents the initial and final reports combined, is being submitted based on the findings of that retrospective review.In accordance with 21 cfr 803.56, if additional information is obtained which was not known or was not available when this report was submitted, then a supplemental report will be submitted to the fda.
 
Event or Problem Description
Lap prostate- "surgeon complaining of camera drag through c0r95 port.Drag noted when moving scope into and out of the port.Camera is olympus endoeye 10mm.Scope cleaning solution used also." patient status unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C0R95, 12X100 KII BLUNT TIP SM CONE 6/BX
Common Device Name
GCJ
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key6372807
Report Number2027111-2017-01049
Device Sequence Number13603863
Product Code GCJ
UDI-Device Identifier00607915127602
UDI-Public(01)00607915127602(17)161027(30)01(10)1208169
Combination Product (Y/N)N
Initial Reporter CountryGB
PMA/510(K) Number
2
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/27/2016
Device Model Number2
Device Catalogue Number2
Device Lot Number1208169
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2014
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2014
Initial Report FDA Received Date03/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS ENDOEYE 10MM
-
-